Model Number MMT-1760KPK
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code Hypoglycemia (1912)
Date of Event 06/07/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Information received by medtronic indicated that the insulin pump had a blank display.The customer was assisted with troubleshooting and display did not returned back.The customer was also reported that battery compartment was neither damaged nor corroded, battery cap was neither missing nor damaged, and the spring was neither damaged nor corroded.The customer was reported that the insulin pump display did not return after restart.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device received with blank display due to corroded electronic assemblies and corroded battery tube.Unable to perform displacement test, rewind, prime/seating, basic occlusion test, force sensor test, occlusion test, self test, and current measurements due to display anomaly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT MMT-1760KPK 670G PATHWAY BLACK MG
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key10153086
Report Number2032227-2020-130810
Device Sequence Number19393239
Product Code OZP
UDI-Device Identifier00763000283445
UDI-Public(01)00763000283445
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-1760KPK
Device Catalogue NumberMMT-1760KPK
Device Lot NumberHG42CTN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2020
Initial Date Received by Manufacturer 06/08/2020
Supplement Date Received by Manufacturer10/06/2020
Initial Report FDA Received Date06/15/2020
Supplement Report FDA Received Date10/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-