|
Review of the device history records indicate that devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.The following devices were also listed in this report: scorpio nrg cr femoral component right #11; 80-4411r ; mhan16 series 7000 standard tibia;7115-0009; mhedyy it cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.Device not returned.
|
|
Reported event: an event regarding infection involving scorpio insert was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned.Clinician review: a review of the provided medical records and/or x-rays by a clinical consultant indicated: undated x-ray ap and lat right knee - cemented tka no patellar resurfacing, reduced, nominal position.(b)(6) 2011 op note "explantation right tka and spacer implantation" diagnosis infection status post right tka gen anesthesia/ tourniquet.".Coag.Negative staph.Crp elevated.Cement spacer." uncomplicated surgery.(b)(6) 2012 op note "knee prosthesis exchange implantation right.Scorpio nrg ps #11 right femur, #11 tibia, 11/12 insert, 2 units palomed cement.Uncomplicated surgery.No clinical or pmh, no dated serial x-rays, no exam of explanted components, no bacteriology lab reports.Based upon the information available for review, confirmation of the event description or preparation of a medical report is not possible." device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there has been no other similar events for the reported sterile lot.Conclusion: all stryker products sold as sterile are validated to a minimum sterility assurance level (sal) of 10^-6 in accordance to applicable iso standards.The exact cause of the event could not be determined because insufficient information was provided.Additional information including pathology reports, operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
|