MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE NANOPASS REACH CRESCENT; PASSER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE NANOPASS REACH CRESCENT; PASSER Back to Search Results
Model Number CAT02298
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/14/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Alleged failure: dr.(b)(6) bent the first one slightly.And was unable to trigger it afterwards.He did not offer the second one, but still it could not be triggered.The third one worked.Procedure could be completed successfully.The failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be: excessive force applied on device, user technique related factors, or difficult anatomy, extremely tough soft tissue.The product was returned for investigation and the failure mode will be monitored for future reoccurrence.No manufacturing date - the device manufacturer date is not known.
 
Event or Problem Description
Per investigation results, the device was broken at the internal shaft.Per additional information received, the device bent inside the joint and when removed was attempted to be straightened and it broke.No pieces broke inside of the joint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NANOPASS REACH CRESCENT
Common Device Name
PASSER
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key10153914
Report Number0002936485-2020-00235
Device Sequence Number2152523
Product Code HWQ
UDI-Device Identifier07613252633204
UDI-Public07613252633204
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCAT02298
Device Catalogue NumberCAT02298
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/29/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/20/2020
Initial Report FDA Received Date06/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-