| Model Number |
X SERIES |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device was unable to detect the attached electrode pads.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical united kingdom; the customer's report was not replicated or confirmed.The device was put through extensive testing including bench handling, power cycling, and full functional stress testing without duplicating the report.The monitor board was replaced as a precaution.The device was recertified and returned to the customer.No trend is associated with reports of this type.Review of the device activity logs did not show any faults that could be associated with customer report.
|
| |
|