Medtronic received information regarding a navigation system being used in a functional endoscopic sinus surgery (fess).It was reported that the power on the system was turned on, but there was an recognition failure in the magnetic field due to an emitter error.It was recognized finally after restarting three times.The same issue recurred even though the product was just recently replaced with a new one due to the same issue.There was less than an hour delay in the case.No impact on patient outcome.Additional information received nothing that there was a communication failure with the em control box.The usb port of the em control box was cleaned with a cotton swab moistened with alcohol.After that, the issue resolved.
|