| Model Number |
MMT-723NAS |
| Medical Device Problem Codes |
Mechanical Problem (1384); Device Difficult to Program or Calibrate (1496); Obstruction of Flow (2423); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/05/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had act button was harder to push.The customer also stated that the insulin pump had lost settings when changing out batteries, rewind was slower, the pump had no delivery alarm.The customer also stated that the battery cap had crack around it.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with all buttons responding properly.No damage on keypad assembly.No unlocked j2/lcd keypad connector.Device passed the displacement test, self test rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test.No unexpected rewind anomalies noted.No excessive no delivery/occlusion alarm noted.Device received with cracked battery tube threads and cracked reservoir tube lip.The test infusion set and reservoir does lock into place.(b)(4).
|
| |
|