MAUDE Adverse Event Report: SMITH & NEPHEW, INC. FLEXIBLE REAMER SHAFT 50CM; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. FLEXIBLE REAMER SHAFT 50CM; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71631192
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/18/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during surgery the flexible reamer shaft 50cm broke at modified hudson junction while using internal to the patient.All pieces were recovered.The procedure was finished using a smith and nephew back up device, with no surgical delay and no injuries to the patient.
 
Additional Manufacturer Narrative
The device, used in treatment, was returned for evaluation.The clinical/medical team concluded, per report all of the pieces were recovered from the patient.No patient injuries or adverse consequences were reported due to the breakage.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.A visual inspection confirmed the tip of the flexible reamer shaft 50cm is cracked.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIBLE REAMER SHAFT 50CM
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10158027
Report Number1020279-2020-02417
Device Sequence Number3867379
Product Code JDG
UDI-Device Identifier03596010419743
UDI-Public03596010419743
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K123598
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71631192
Device Catalogue Number71631192
Device Lot Number15FM02477
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2020
Initial Date Received by Manufacturer 05/18/2020
Supplement Date Received by Manufacturer01/27/2021
Initial Report FDA Received Date06/16/2020
Supplement Report FDA Received Date02/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-