| Model Number |
601184 |
| Medical Device Problem Codes |
Application Program Problem: Parameter Calculation Error (1449); Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical inc.Internal report number (b)(4).The b.Braun pinnacle support technician provided instructions to perform the factory calibration.Customer was advised that under no circumstances should any user ever circumvent the calibration process as this places patient safety at risk.After performing the factory calibration the 1000 gram calibration weight read 1000 grams on the load cell.The device involved has not been received for evaluation and the investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
As per reported by the user facility: customer reported the 1000 gram weight measured 983 grams during calibration.They placed additional weights on the load cell until the calibration read the scale at 1 kilogram.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical inc.Internal report number (b)(4).The b.Braun pinnacle support technician provided instructions to perform the factory calibration.Customer was advised that under no circumstances should any user ever circumvent the calibration process as this places patient safety at risk.After performing the factory calibration the 1000 gram calibration weight read 1000 grams on the load cell.The device involved was not returned for evaluation.If additional pertinent information becomes available a follow-up report will be filed.
|
| |
|
Search Alerts/Recalls
|