| Catalog Number |
0035280 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/25/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that foreign material was found inside the syringe.
|
| |
|
Event or Problem Description
|
|
It was reported that the foreign material found inside the syringe.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was confirmed, however the cause was unknown.Visual inspection noted that one irrigation syringe was received without the original packaging.Visual evaluation noted that 3 white specs (measuring 0.20, 0.25, 0.20 sq.Mm) were found in the bulb outside the syringe body.Although the reported event was confirmed, the root cause could not be determined.A potential root cause for this failure mode could be due to defective/ contaminated components from suppliers.The product was not used for diagnosis or treatment.The product had failed to meet the specifications, and was influenced by the reported failure.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The labelling review was not performed due to labeling could not have prevented the reported failure.
|
| |
|