DEPUY SYNTHES PRODUCTS LLC HUDSON REAMING ATTACHMENT F/BPL; MOTOR, SURGICAL INSTRUMENT, AC-POWERED
|
Back to Search Results |
|
| Catalog Number |
530.782 |
| Medical Device Problem Code |
Excessive Heating (4030)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/10/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Concomitant med products and therapy dates: (2) battery reciprocator devices, battery reamer/drill device, battery oscillator device, chuck with key device, reaming attachment device, battery device; (b)(6) 2020.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi ¿ (b)(4).
|
| |
|
Event or Problem Description
|
|
This is report 6 of 7 for the same event.It was reported from (b)(6) that during an unspecified surgical procedure it was observed that two battery reciprocator devices, a battery reamer/drill device, a battery oscillator device, a chuck with key device, two reaming attachment devices, and a battery device were making abnormal noise and generating heat.It was not reported if there were any delays in the procedure due to the event.It was not reported if there was a spare device available for use.There was patient involvement reported.There were no reports of any injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The actual device was returned for evaluation.The reaming attachment device was evaluated and the reported condition that the device was making an abnormal noise and generating heat was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was observed that the device had component damage.The assignable root cause of this condition was determined to be traced to component failure due to normal wear.
|
| |
|
Search Alerts/Recalls
|
|
|