ARTHREX, INC. APS II SHVR HANDP,FT-SWCH CNTRL; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
Model Number APS II SHVR HANDP,FT-SWCH CNTRL |
Device Problem
Defective Device (2588)
|
Patient Problem
No Code Available (3191)
|
Event Date 03/10/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.
|
|
Event Description
|
It was reported that during a procedure, the motor on the apsii shaver, ar-8330f, was not working and the surgeon had to switch to an open procedure.Additional information requested.Additional information provided 6/1/2020: this occurred during an acl procedure and the shaver did not show any alarm or error messages.
|
|
Manufacturer Narrative
|
(motor hall sensor) the evaluation confirmed the reported event.The cause was attributed to a defective motor hall sensor.
|
|
Search Alerts/Recalls
|
|
|