SMITH & NEPHEW, INC. EVOS SMALL T8 TAPERED DR SHAFT - SHORT; SCREWDRIVER Back to Search Results
Model Number 71175078
Medical Device Problem Code Defective Device (2588)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/23/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery the short t8 power driver head twisted while inserting screw (inside the patient).The procedure was completed without delay.Backup from smith&nephew was available.No patient injury or other complications were reported at the moment.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
E1: information was corrected.
 
Additional Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection confirmed the tip is damaged on the evos small t8 tapered dr shaft - short.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOS SMALL T8 TAPERED DR SHAFT - SHORT
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10159971
Report Number1020279-2020-02426
Device Sequence Number3867383
Product Code HXX
UDI-Device Identifier00885556636084
UDI-Public00885556636084
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71175078
Device Catalogue Number71175078
Device Lot Number17MUP0005
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 05/23/2020
Supplement Date Received by Manufacturer05/23/2020
08/07/2020
Initial Report FDA Received Date06/16/2020
Supplement Report FDA Received Date06/18/2020
08/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-