According to the reporter, following a ventral hernia repair in 2018, the patient had felt pain which was similar to the original hernia.The patient underwent laparoscopic exploration for the pain and it was found that there were adhesions to the mesh.No mesh was in the removed adhesion but dissection of the mesh revealed two holes in the center of the mesh.Additional mesh was placed over the hole in the existing mesh.The surgical time was extended and the patient was anesthetized for 40 additional minutes.
|
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.It was reported that the device torn.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: the possible complications associated with the use of symbotex¿ composite mesh are those typically associated with surgically implantable mesh: seroma, hematoma, recurrence, adhesions, fistula formation, infection, inflammation, chronic pain, and/or allergic reactions to the components of the product.Potential events associated with state of the art mesh with similar indication may include: organ injury (including bowel and visceral injury), trocar-site herniation, bowel obstruction and urinary retention (related with the use of anesthetics).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|