Model Number SYM2015F
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Codes Adhesion(s) (1695); Pain (1994); No Code Available (3191); Unspecified Tissue Injury (4559)
Date of Event 06/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, following a ventral hernia repair in 2018, the patient had felt pain which was similar to the original hernia.The patient underwent laparoscopic exploration for the pain and it was found that there were adhesions to the mesh.No mesh was in the removed adhesion but dissection of the mesh revealed two holes in the center of the mesh.Additional mesh was placed over the hole in the existing mesh.The surgical time was extended and the patient was anesthetized for 40 additional minutes.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.It was reported that the device torn.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: the possible complications associated with the use of symbotex¿ composite mesh are those typically associated with surgically implantable mesh: seroma, hematoma, recurrence, adhesions, fistula formation, infection, inflammation, chronic pain, and/or allergic reactions to the components of the product.Potential events associated with state of the art mesh with similar indication may include: organ injury (including bowel and visceral injury), trocar-site herniation, bowel obstruction and urinary retention (related with the use of anesthetics).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYMBOTEX
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR  01600
MDR Report Key10161338
Report Number9615742-2020-01304
Device Sequence Number11490804
Product Code FTL
UDI-Device Identifier10884521190665
UDI-Public10884521190665
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K131969
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2022
Device Model NumberSYM2015F
Device Catalogue NumberSYM2015F
Device Lot NumberPRC0101X
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/01/2020
Supplement Date Received by Manufacturer06/25/2020
Initial Report FDA Received Date06/17/2020
Supplement Report FDA Received Date12/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
-
-