| Model Number |
TSVM6B11 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
05/26/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint: (b)(4).Patient's weight is unknown.Doctor's first name was not provided.Additional 510(k) number is k101880.A summary investigation has been completed for lack of primary stability events that do not allege a deficiency with the implant and identified that the reported event is likely due to biological factors which have an adverse effect on implant stability or is related to surgical technique and insertion torque.Due to a wide range of external factors (non-design/ non-manufacturing related), identifying a definitive root cause is generally not possible.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the implant was unable to be placed due to the lack of primary stability.Patient was bone grafted and sent home for future implant placement.Tooth number is unknown.
|
| |
|