• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number MT22495-BLU
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that unexpected receiver shutdown occurred.No product or data was provided for evaluation.Confirmation of the allegation and a probable cause could not be determined.No injury or medical intervention was reported.
 
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The product was evaluated.An external visual inspection was performed and passed.Receiver charge and boot tests were performed and failed due to the receiver not powering on.Functional tests were not performed due to the receiver not powering on.An internal visual inspection was performed and failed due to a damaged battery.Pictures were taken.The receiver log was downloaded using a good known battery.The receiver log was reviewed finding error related to the complaint.The allegation was confirmed.The probable cause was determined to be receiver damaged battery.No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G4 PLATINUM CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
MDR Report Key10162716
Report Number3004753838-2020-059092
Device Sequence Number19503171
Product Code MDS
UDI-Device Identifier30386270000171
UDI-Public30386270000171
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P120005
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMT22495-BLU
Device Catalogue NumberSTR-DR-BLU
Device Lot Number5224945
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/02/2020
Supplement Date Received by Manufacturer08/16/2020
Initial Report FDA Received Date06/17/2020
Supplement Report FDA Received Date08/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age72 YR
Patient Weight61
-
-