• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: US SURGICAL PUERTO RICO ILA; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

US SURGICAL PUERTO RICO ILA; STAPLE, IMPLANTABLE Back to Search Results
Model Number 3948L
Medical Device Problem Code Failure to Fire (2610)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during colon surgery, when on a functional side-to-side anastomosis, the device was unable to be fired.When a new cartridge was attached, it was able to be fired without problem.There was no patient injury.
 
Additional Manufacturer Narrative
Additional information: d3(mfr name, street 1, mfr city, mfr region, postal code), d10, g4, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that the device did not fire.The reported issue could not be confirmed.The most likely root cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are also thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ILA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce,pr CT 00716 4401
MDR Report Key10163271
Report Number1219930-2020-02412
Device Sequence Number3514297
Product Code GDW
UDI-Device Identifier10884521063150
UDI-Public10884521063150
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K885047
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model Number3948L
Device Catalogue Number3948L
Device Lot NumberP9D1504Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2020
Supplement Date Received by Manufacturer07/14/2020
Initial Report FDA Received Date06/17/2020
Supplement Report FDA Received Date07/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-