• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of this report: 17 jun 2020.
 
Event Description
It was reported there was an oxygen supply fault during continuous built in testing.There was no patient involvement.
 
Manufacturer Narrative
G4: 06oct2020.The remote service engineer advised the customer to change the o2 percentage to different levels during testing and monitor the o2 delivery with an external sensor.The customer confirmed one occurrence of the oxygen supply fault error in the significant error log.The customer ran the high and low pressure leak tests which passed and checked o2 levels at different settings with v60 running for over half a day without error.No repairs were made as the error was resolved through testing and the device was returned to use.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key10165056
MDR Text Key196065659
Report Number2031642-2020-02114
Device Sequence Number1
Product Code MNT
UDI-Device Identifier0884838009851
UDI-Public(01)0884838009851
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 06/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/17/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-