| Model Number |
60ENY |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The incident sample has been requested but to date has not been received for evaluation. if the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted. as part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
|
| |
|
Event or Problem Description
|
|
The customer reported that the nurse could not flush feeding through extension tubing.There was no patient harm reported.The extension tubing was replaced.The nasogastric (ng) tube did not require a replacement.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record of the reported lot number shows evidence that the product was released according to all established procedures and qa documentation one decontaminated sample was received for evaluation.The reported condition is confirmed.The most likely root cause for this condition could be a result of a workmanship issue.The manufacturing process was reviewed and during the assembly process it is possible for excess solvent to be applied to the device during the bonding process generated by dispenser when the application is performed without a guide pin.The reported condition was evaluated against the acceptable quality limit (aql) and no action plan is required since the complaint report represents an occurrence rate of 0.003% which is below the approved aql of 0.65 %.The current process is running according to product specifications meeting quality acceptance criteria.We will continue monitoring the process for any adverse trends that require immediate attention.
|
| |
|
Search Alerts/Recalls
|