It was reported that during thr revision surgery (it is unknown if was a revision surgery of smith and nephew devices), while using the drill to drill through a cement mantle in the femur, the tip of the cement drill broke off.The procedure was finished with a change in the surgical technique.There was no surgical delay and no injury to the patient.
|
|
H3,h6: the device was not returned for evaluation and the reported event could not be confirmed.The contribution of the device to the reported event could not be corroborated.The clinical/medical investigation concluded that, a cement drill broke during a hip revision.¿the procedure was finished with a change in the surgical technique.There was no surgical delay and no injury to the patient.¿ the impact to the patient beyond the surgery and recovery cannot be determined.Smith and nephew has not received adequate materials to fully evaluate the complaint, but if additional clinically relevant materials are later received, then the case may be re-opened for further evaluation.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Additional information: d3.
|