PLEXUS MANUFACTURING SDN. BHD MYCARELINK PATIENT MONITOR; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number 24950LLQ
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/02/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the remote monitor was too hot to touch.No troubleshooting was taken.The remote monitor is expected to be rep laced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product analysis: model#: 24950llq, serial/lot#: (b)(6): the remote monitor was returned and analysis revealed that the remote monitor failed during the interrogation test.In addition, the radiofrequency head battery was defective and was returned with gold contact.Lastly, error code 3230 was found in the failure analysis log.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The remote monitor was returned and replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYCARELINK PATIENT MONITOR
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
MDR Report Key10166373
Report Number3004593495-2020-00629
Device Sequence Number2142784
Product Code NVZ
UDI-Device Identifier00763000113995
UDI-Public00763000113995
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P890003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number24950LLQ
Device Catalogue Number24950LLQ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2020
Supplement Date Received by Manufacturer08/07/2020
Initial Report FDA Received Date06/18/2020
Supplement Report FDA Received Date08/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age73 YR
-
-