The optional handheld barcode scanner used for the dca vantage analyzer allows scanning of operator id, patient id, sample demographics, and comments.Barcodes with a check-digit have an additional digit at the end of the unique alpha/numeric string, providing an extra functionality that should reduce the frequency of erroneous scans.The scanner is not confirming the integrity of read barcode data using the check-digit.As a result there are three issues identified with the reading of code 39 barcodes using handheld barcode scanner ds4308: the scanner might allow misinterpretation of barcode data which may lead to incorrect data entry for sample id, patient id and comments because it matches an existing sample id, patient id or comments value.This can cause the results of a particular sample to be mis-identified and potentially assigned to a different patient.In order for this unlikely event to occur, incorrect decoding of the barcode by the scanner must occur, and the incorrect sample id, patient id and comments must match a different existing sample id, patient id or comments.The scanner might allow misinterpretation of barcode data which may lead to incorrect data entry for sample id, patient id and comments that does not match any existing sample id, patient id or comments.This may lead to the sample being unmatched until the correct patient id is entered.The scanner might allow misinterpretation of barcode data which may lead to incorrect data entry for operator id.If this occurs two effects are possible.It may prevent the user from being immediately authenticated.This issue can be mitigated by rescanning the barcode.If the incorrect operator id matches a different existing operator id with added permissions the operator could have access to additional dca functions.This issue was found internally.There are no reported customer complaints.
|