H10: only (1) actual sample was received for evaluation.Visual inspection was performed which observed a small cut on the tubing.A functional testing was performed which revealed a leak on the affected area.The reported condition was verified.The cause of the condition was due to the set left protruded while loading it into the blister cavity which resulted in a cut by the blade machine during the packaging step of the manufacturing process.The other sample was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|