| Model Number |
9-PDA-003 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/25/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Further information regarding this event has been requested.The investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2020 an amplatzer duct occluder that was implanted was discovered to have a bulbous deformation just after implant.It was removed and replaced with a competitor's product.Patient remained hemodynamically stable throughout the procedure and continues to remain stable.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of device deformation could not be confirmed.The investigation confirmed the device met visual and functional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of an amplatzer duct occluder deploying with a bulbous deformation could not be confirmed.A more comprehensive assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: d2.
|
| |
|