Model Number OER-PRO
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/26/2020
Event Type  malfunction  
Manufacturer Narrative
Based on the on-site evaluation, the field service engineer observed the detergent nozzle was leaking water.Cleaned and tightened detergent nozzle.During repair, found that the front door was not shutting properly and replaced door latch.Noted that the push plate was cracked.Replaced the push plate.While removing the lid to replace the push plate, found that the lid was cracked.Replaced the lid.Equipment repaired and verified according to oem instructions.Software attributes have been verified and confirmed.Equipment passed the electrical safety test and results.The cause could be traced to maintenance.No further information was reported.
 
Event Description
The customer reported to olympus there was fluid leakage inside the device.There was no patient injury reported.
 
Manufacturer Narrative
The investigation has been completed.A device history record (dhr) review was performed and found no non-conformances that would cause the reported malfunction.All records indicate that the device was manufactured in accordance with all applicable procedures.The instructions for use (ifu) states: ¿before using the equipment, always check that there is no irregularity regarding the following points on the lid and the lid packing.If there is any irregularity, cleaning fluid or disinfectant solution may leak out.¿ the root cause could not be conclusively determined.Probable causes that could have led to the reported event include due to the user handling such as the user dropped an object on the lid.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPE REPROCESSOR
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10171910
MDR Text Key206391671
Report Number8010047-2020-03461
Device Sequence Number1
Product Code FEB
UDI-Device Identifier04953170258589
UDI-Public04953170258589
Combination Product (y/n)N
PMA/PMN Number
K103264
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 12/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-PRO
Was the Report Sent to FDA? No
Date Manufacturer Received11/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-