• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON ASYMMETRIC X3 PATELLA; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON ASYMMETRIC X3 PATELLA; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5551-G-350
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Joint Dislocation (2374); Ambulation or Postural Difficulties (2544)
Date of Event 09/02/2014
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Information received includes a report dated (b)(6) 2014: ".Rectus femoris complete rupture and retracted 8cm from the superior pole of the patella.The quadriceps tendon will be repaired.".
 
Event or Problem Description
Information received includes a report dated september 2, 2014: "rectus femoris complete rupture and retracted 8cm from the superior pole of the patella.The quadriceps tendon will be repaired.".
 
Additional Manufacturer Narrative
Reported event: an event regarding patient factors involving a triathlon insert was reported.The event of aspiration was confirmed by medical review.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remained implanted.Clinical review: review of the provided medical records by a clinical consultant indicated: revision surgery was performed on (b)(6) 2013.The preoperative diagnosis was now failed knee arthroplasty with femoral component loosening and polyethylene liner wear.At the time of the revision the patellar component was found completely floating in the knee.The femoral and tibial components were stable and well aligned.The patella was recut and a a3 5 x 10mm patella cemented in place.A new 9 mm cs polyethylene was placed.A lateral release and medial imbrication performed as well.No complications were reported from the surgery.Postoperative x-rays showed satisfactory appearance of the implants.An mri was reviewed, but it is unclear if this was the same mri that was referenced in the on (b)(6) visit.The mri was felt to show rectus femoris complete rupture and retraction 8 cm.There was dimpling at the skin, 3/5 strength with straight leg raise, and some effusion.A plan was made for repair of the quad with "internal bracing" and prp injection.Operative notes, hospital records and immediate postoperative records were not provided.The patient was doing well, was flexing to 50°, and could walk without her brace with a walker without a limp.Hazards: none clearly related to implant.Conclusion of assessment: the revision surgery was confirmed as were the two quadriceps repairs.Product history review: indicated all devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been 1 other similar events for the reported lot related to the same device / patient.Conclusions: review of the provided medical records by a clinical consultant indicated: he mri was felt to show rectus femoris complete rupture and retraction 8 cm.There was dimpling at the skin, 3/5 strength with straight leg raise, and some effusion.A plan was made for repair of the quad with "internal bracing" and prp injection.Hazards: none clearly related to implant.Conclusion of assessment: the revision surgery was confirmed as were the two quadriceps repairs.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON ASYMMETRIC X3 PATELLA
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key10173124
Report Number0002249697-2020-01236
Device Sequence Number14331297
Product Code MBH
UDI-Device Identifier07613327014808
UDI-Public07613327014808
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/30/2018
Device Model Number5551-G-350
Device Catalogue Number5551-G-350
Device Lot Number035W
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/02/2020
Supplement Date Received by Manufacturer08/14/2020
Initial Report FDA Received Date06/19/2020
Supplement Report FDA Received Date09/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age75 YR
Patient Weight104
-
-