• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS LTD ACETABULAR CUP HAP SIZE 40/46; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS LTD ACETABULAR CUP HAP SIZE 40/46; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING Back to Search Results
Catalog Number 74122146
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Test Result (2695); No Code Available (3191); Joint Laxity (4526); Metal Related Pathology (4530)
Date of Event 06/23/2014
Type of Reportable Event Serious Injury
Event or Problem Description
Can legal it was reported that a revision surgery was performed on the patients left hip due to metallosis.The patient outcome is unknown.
 
Additional Manufacturer Narrative
It was reported that left hip revision surgery was performed.During the revision, the hemi head and sleeve were removed.The bhr cup and stem remained implanted.As of today, the implanted devices, all of which were used in treatment, and additional information have been requested for this complaint but have not become available.A review of the complaint history for the bhr cup, modular head and modular sleeve was performed using batch numbers in search of similar recurring reports for the products during their lifetimes.No other similar complaints were identified for the bhr cup.Similar complaints have been identified for the modular head and sleeve.However, as the devices are no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.Review of manufacturing records did not reveal any waivers, concessions, manufacturing or material abnormalities that could have contributed to this issue.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.No additional risks were identified as result of the reported event and no further actions are required at this time.The available medical documents were reviewed.Although acetabular loosening was noted pre-revision, no medical imaging was provided and no acetabular loosening was noted intraoperatively.Without the supporting lab/pathology results, imaging, and/or the analysis of the explanted components, the root cause of the reported acetabular loosening and noted intraoperative findings of metallosis cannot be confirmed, and it cannot be concluded that the reported reactions were associated with a mal-performance of the implant.The patient impact beyond the revision, and expected transient post-op convalescence period cannot be determined.Without return of the actual devices or further information we cannot further investigate or confirm the details supplied in this complaint, and our investigation remains inconclusive.If the products or additional information become available in the future, this case will be reopened.No preventative or corrective action has been initiated as a result of this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACETABULAR CUP HAP SIZE 40/46
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
MDR Report Key10173274
Report Number3005975929-2020-00197
Device Sequence Number2974141
Product Code NXT
UDI-Device Identifier03596010565792
UDI-Public3596010565792
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P040033
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Device Explanted Year2014
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/29/2012
Device Catalogue Number74122146
Device Lot Number07EW11840
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/22/2020
Supplement Date Received by Manufacturer12/27/2020
Initial Report FDA Received Date06/19/2020
Supplement Report FDA Received Date12/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-