| Model Number |
MMT-551NAB |
| Medical Device Problem Codes |
Failure to Cycle (1142); Detachment of Device or Device Component (2907); Priming Problem (4040)
|
| Health Effect - Clinical Code |
Hyperglycemia (1905)
|
| Date of Event |
06/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had compromised force sensor system.Customer stated that drive support cap was sticking out.Customer was priming the insulin pump.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed displacement test.However device alarmed motor error during basic occlusion test due to loose or protruded drive support disk noted.Unable to perform occlusion test, prime or a33 test and excessive no delivery or verify a33 alarm and prime or fill anomaly due to motor error.
|
| |
|