• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. COMP AUG MINI BSPLT W TPR LG; PROSTHESIS, EXTREMITIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. COMP AUG MINI BSPLT W TPR LG; PROSTHESIS, EXTREMITIES Back to Search Results
Catalog Number 110032430
Medical Device Problem Codes Loosening of Implant Not Related to Bone-Ingrowth (4002); Migration (4003)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 05/28/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Concomitant medical products: part#113632 lot#349340.Part#110031399 lot#64538350.Part#110031427 lot#64336651.Part#110030776 lot#11024227.Part#115395 lot#783220.Part#180551 lot#035300.Part#180553 lot#599050.Part#180552 lot#641230.Part#180552 lot#819300.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as it is not available to be returned per hospital policy.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that patient presented for 6 month follow up post right shoulder hemiarthroplasty where x-rays revealed a broken central screw with glenoid loosening and migration.The patient underwent a revision surgery approximately 6 days later.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The reported products were reviewed for compatibility with no issues noted.Medical records/radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: 22-may-2020; 6 months postop.Glenoid component migration of 5mm, glenoid implant fracture with loosening noted on xray.Rom: elevation 142°, 20-70°, t12.Moderate instability (3/10), mild pain (2-10), ases 70.7, health score 100.Ae: central screw breakage with shift in glenoid component, date of onset 22-may-2020, planned resection and vrs implant occurred on 28-may-2020.Dhr was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP AUG MINI BSPLT W TPR LG
Common Device Name
PROSTHESIS, EXTREMITIES
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10173939
Report Number0001822565-2020-02131
Device Sequence Number16914767
Product Code KWS
UDI-Device Identifier00880304993259
UDI-Public(01)00880304993259
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K172502
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number110032430
Device Lot Number64376398
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/26/2020
Supplement Date Received by Manufacturer10/05/2020
Initial Report FDA Received Date06/19/2020
Supplement Report FDA Received Date10/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Weight108
-
-