• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. COMP RVS CNTRL 6.5X25MM ST/RST; PROSTHESIS, EXTREMITIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. COMP RVS CNTRL 6.5X25MM ST/RST; PROSTHESIS, EXTREMITIES Back to Search Results
Catalog Number 115395
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 05/28/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2020-02423, 0001825034-2020-02427, 0001825034-2020-02428, 0001825034-2020-02429.Concomitant medical products: part#113632, lot#349340.Part#110031399, lot#64538350.Part#110031427, lot#64336651.Part#110030776, lot#11024227 part#110032430, lot#64376398.Part#180551, lot#035300.Part#180553, lot#599050.Part#180552, lot#641230.Part#180552, lot#819300.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as it is not available to be returned per hospital policy.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that patient presented for 6 month follow up post right shoulder hemiarthroplasty where x-rays revealed a broken central screw with glenoid loosening and migration.The patient underwent a revision surgery approximately 6 days later.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g4; g7; h1; h2; h3; h6.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The reported products were reviewed for compatibility with no issues noted.Medical records/radiographs were provided and identified the following, baseplate failure and the screws being fractured.Device history record was reviewed and no discrepancies relevant to the eported event were found.A definitive root cause cannot be identified.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP RVS CNTRL 6.5X25MM ST/RST
Common Device Name
PROSTHESIS, EXTREMITIES
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10174077
Report Number0001825034-2020-02363
Device Sequence Number16585015
Product Code KWS
UDI-Device Identifier00880304677074
UDI-Public(01)00880304677074
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K132239
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number115395
Device Lot Number783220
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/26/2020
Supplement Date Received by Manufacturer09/03/2020
Initial Report FDA Received Date06/19/2020
Supplement Report FDA Received Date09/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
NI
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NARRATIVE IN H10
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Weight108
-
-