| Model Number |
MMT-723NAS |
| Medical Device Problem Codes |
Mechanical Problem (1384); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
04/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had motor error alarm.Customer reported that the drive support cap was sticking out.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Search Alerts/Recalls
|