• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDICAL COMPONENTS, INC. 4F SINGLE CT MIDLINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDICAL COMPONENTS, INC. 4F SINGLE CT MIDLINE Back to Search Results
Model Number MLCT4S20
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/03/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is ongoing.A follow up report will be submitted when the investigation is complete.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Guidewire tip knotted and broke into four (4) pieces which required surgery for removal.
 
Additional Manufacturer Narrative
No device was returned and no photographs were provided.A review of the manufacture records was conducted by the guidewire contract manufacturer and revealed the device was manufactured according to specification with no non-conformances or deviations in place.Without an evaluation of the complained device a definitive root cause cannot be determined.The instructions for use contain the following caution statements: do not advance the guidewire or catheter if unusual resistance is encountered.Do not insert or withdraw the guidewire forcibly from any component.The wire may break or unravel.If the guidewire becomes damaged, the introducer needle or sheath/dilator and guidewire must be removed together.Device was used for treatment, not diagnosis.If information is obtained that was not avail able for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4F SINGLE CT MIDLINE
Common Device Name
CT MIDLINE
Manufacturer (Section D)
MEDICAL COMPONENTS, INC.
1499 delp drive
harleysville PA 19438
MDR Report Key10174159
Report Number2518902-2020-00024
Device Sequence Number2680696
Product Code PND
UDI-Device Identifier00884908113730
UDI-Public884908113730
Combination Product (Y/N)N
Initial Reporter CountryIT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date02/10/2023
Device Model NumberMLCT4S20
Device Catalogue NumberMLCT4S20
Device Lot NumberMNXA650
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/10/2020
Supplement Date Received by Manufacturer06/10/2020
Initial Report FDA Received Date06/19/2020
Supplement Report FDA Received Date07/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-