| Model Number |
MMT-523LNAP |
| Medical Device Problem Codes |
Device Difficult to Program or Calibrate (1496); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the act or escape keys were not responding.Customer stated that the time was advancing.Customer was unable to remove battery cap.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with intermittent button response due to flattened express bolus, esc, act, up and down button dome switch (creased).No button error alarm during testing.No unlocked j2/lcd keypad connector.Device passed displacement test and self test.Device received with stripped battery cap coin slot(p).
|
| |
|