• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR MITRACLIP SYSTEM STEERABLE GUIDE CATHETER; VALVE REPAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MITRACLIP SYSTEM STEERABLE GUIDE CATHETER; VALVE REPAIR Back to Search Results
Model Number SGC0301
Device Problem Leak/Splash (1354)
Patient Problem No Patient Involvement (2645)
Event Date 06/02/2020
Event Type  malfunction  
Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report the leak.It was reported that this was a mitraclip procedure to treat grade 4 functional mitral regurgitation (mr).During device preparation prior to use, the dilator from the steerable guide catheter (sgc) was leaking fluid from the rotating hemostatic valve when it was flushed.The connection was tightened, but the leak continued.This dilator was not used for the procedure.Another sgc was used to complete the procedure.No additional information was provided.
 
Manufacturer Narrative
All available information was investigated and the reported issue (leak) was confirmed via returned device analysis.Although a review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event, the investigation determined the reported hemostasis valve leak appears to be related to a potential product quality issue.The issue is being addressed per internal operating procedures.Abbott vascular (av) will continue to trend the performance of these devices.Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of Device
VALVE REPAIR
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10174435
MDR Text Key195891620
Report Number2024168-2020-05175
Device Sequence Number1
Product Code DRA
UDI-Device Identifier08717648216824
UDI-Public08717648216824
Combination Product (y/n)N
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Remedial Action Notification
Type of Report Initial,Followup
Report Date 08/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/19/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/10/2021
Device Model NumberSGC0301
Device Catalogue NumberSGC0301
Device Lot Number00110U138
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/16/2020
Date Manufacturer Received07/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-