• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. PERCUTANEOUS GUIDE BOLT; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. PERCUTANEOUS GUIDE BOLT; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71631024
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Injury (2348)
Date of Event 05/26/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during surgery when was insert the nail, the guide came off the nail when impacting, this was because the tip of the guide bold broke off in the nail inside the patient.The guide was removed and the piece in the nail had to be fished out.The second guide bolt was used and the nail was inserted.The procedure had a delay between 0-30 min.Backup from smith&nephew was available.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.So the reported event could not be confirmed.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERCUTANEOUS GUIDE BOLT
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10174774
Report Number1020279-2020-02610
Device Sequence Number2153538
Product Code HSB
UDI-Device Identifier03596010604903
UDI-Public03596010604903
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71631024
Device Catalogue Number71631024
Device Lot Number11FM15164
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/27/2020
Supplement Date Received by Manufacturer05/25/2021
Initial Report FDA Received Date06/19/2020
Supplement Report FDA Received Date05/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-