MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 DXTEND MOD EPI 1 ECC RIGHT HA; DELTA XTEND IMPLANTS : SHOULDER HUMERAL STEM ACCESSORY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 DXTEND MOD EPI 1 ECC RIGHT HA; DELTA XTEND IMPLANTS : SHOULDER HUMERAL STEM ACCESSORY Back to Search Results
Model Number 130720103
Medical Device Problem Code Loss of Osseointegration (2408)
Health Effect - Clinical Codes Unspecified Infection (1930); Inadequate Osseointegration (2646); No Code Available (3191)
Date of Event 06/05/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The patient was revised to address infection and loosening of the stem at the bone to implant interface.The surgeon was planning on doing an i&d of a reverse tsa and once he got into the shoulder joint, the pressfit global unite stem was loose so he took out the stem, modular epiphysis, humeral cup and lateralized glenosphere and implanted a new lateralized glenosphere, cemented monobloc stem and a humeral cup.No height and weight was available at that time.Doi: (b)(6) 2020, dor: (b)(6) 2020, right shoulder.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H6 patient code: no code available (3191) was used to capture device revision or replacement.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND MOD EPI 1 ECC RIGHT HA
Common Device Name
DELTA XTEND IMPLANTS : SHOULDER HUMERAL STEM ACCESSORY
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key10175378
Report Number1818910-2020-14166
Device Sequence Number14336657
Product Code KWS
UDI-Device Identifier10603295027300
UDI-Public10603295027300
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
PMA/510(K) Number
K021478
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number130720103
Device Catalogue Number130720103
Device Lot Number5360155
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/05/2020
Supplement Date Received by Manufacturer06/05/2020
07/21/2020
Initial Report FDA Received Date06/19/2020
Supplement Report FDA Received Date06/23/2020
07/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DXTEND MOD EPI 1 ECC RIGHT HA; DXTEND STAND PE CUP D38 +6MM; GLOBAL UNITE STD STEM SZ 10; XTND GLENO ECC D38MM +2MM; DXTEND MOD EPI 1 ECC RIGHT HA; DXTEND STAND PE CUP D38 +6 MM; GLOBAL UNITE STD STEM SZ 10; XTND GLENO ECC D38 MM +2 MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
-
-