| Model Number |
130720103 |
| Medical Device Problem Code |
Loss of Osseointegration (2408)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Inadequate Osseointegration (2646); No Code Available (3191)
|
| Date of Event |
06/05/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
The patient was revised to address infection and loosening of the stem at the bone to implant interface.The surgeon was planning on doing an i&d of a reverse tsa and once he got into the shoulder joint, the pressfit global unite stem was loose so he took out the stem, modular epiphysis, humeral cup and lateralized glenosphere and implanted a new lateralized glenosphere, cemented monobloc stem and a humeral cup.No height and weight was available at that time.Doi: (b)(6) 2020, dor: (b)(6) 2020, right shoulder.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H6 patient code: no code available (3191) was used to capture device revision or replacement.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.Based on previous investigations, this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.
|
| |
|