The reported event that long nail kit r1.5, ti, right gamma3® ø13x340mm x 125° was alleged of issue ¿user related event - not product related¿ could be confirmed with the help of information provided by the sales representative.A device inspection was not possible since the affected device was not returned, and no other evidences were provided for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Based on the investigation, the root cause of the failure is user related as the user inserted a right nail into a left femur which is an off-label use.It has also been stated in warning and precautions section in ifu as ¿4.2 intra-operative implants can be available in different versions, varying for example in length, diameter, angle, right-hand and left-hand versions, material and number of drilled holes.Select the required version carefully.¿ for the surgeon to choose the required version carefully.If the product is returned or any additional information is provided, the investigation will be reassessed.Device disposition is unknown.
|
|
As reported: (b)(6) hospital was doing a gamma nail and inserted a right nail into a left femur.Due to the bow in the nail, the nail fractured the distal anterior cortex of the pt's femur busting out through the bone.Not knowing what caused this to occur, the surgeon opened a longer nail to try and by-pass this newly created fracture but this too was a right (wrong sided) nail and the same issue occurred with the nail going through the same distal anterior fracture.It was at this time they had realised the wrong sided nails were used both times.It was not long after this i arrived at the hospital.Surgical delay of approx 60 minutes.
|