| Catalog Number |
H93813901 |
| Medical Device Problem Codes |
Delivered as Unsterile Product (1421); Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
05/27/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the packaging of eighty (80) rapidfill connectors were wet.This was identified prior to use.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was added to d10, h3, h4 and h6.H10: three (3) actual samples were received for evaluation.Unaided visual inspection was performed which observed contaminated dark/transparent areas of the primary packaging on the back side and seen through on the front side of all samples.The cause of the condition was not determined.Functional testing was not performed for this complaint.The remaining seventy-seven (77) samples were not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|