| Model Number |
AIRLIFE¿ ADULT MANUAL RESUSCITATOR, OXYGEN RESERVOIR BAG, EXPIRATORY FILTER |
| Medical Device Problem Code |
Loose or Intermittent Connection (1371)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/19/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device or picture has not been returned for evaluation.Therefore, root cause has not been determined.Additionally, a device history record (dhr) of the lot number revealed no issues.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The customer reported that the oxygen tubing got disconnected from the wall (o2 source) when the o2 pressure increased.A gas leak was then observed from the airlife adult manual resuscitator.It happened during patient use, and the insufflator was replaced with a different brand.The incident did not cause any harm but did delay treatment.
|
| |
|
Additional Manufacturer Narrative
|
|
Result of investigation: one sample of the p/n 2k8004f without lot number was received for the investigation.According to investigation, defect reported on this complaint was not confirmed, sample provided meet with all specifications.It is not necessary a root cause for this complaint.
|
| |
|