MEDTRONIC HEART VALVES DIVISION HANCOCK; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number HC105-22
Medical Device Problem Codes Fracture (1260); Gradient Increase (1270); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Pulmonary Insufficiency (2022); Pulmonary Valve Stenosis (2024); No Information (3190)
Date of Event 05/22/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device remains implanted; therefore, no product analysis can be performed.Without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that 9 years and 11 months post implant of this 22mm pulmonary valved conduit in a 9 year old pediatric patient, a transcatheter pulmonary bioprosthetic valve (tpbv) was implanted valve-in-valve.The reason for replacement was not reported.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
Medtronic received additional information that the reason for replacement was due to stenosis, moderate insufficiency, increased gradients of 85mmhg and fracture of the left pulmonary artery stent.The newly implanted valve also had trivial insufficiency the physician thought to be acceptable.No additional adverse patient effects were reported. medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic received additional information that 9 years and 7 months post implant of the pulmonary valved conduit, a balloon valvuloplasty was performed.Then, 9 years and 11 months post implant a transcatheter pulmonary valve (tpv) was implanted due to mild-to-moderate distal stenosis with a gradient of 85mmhg and moderate insufficiency.It was reported that "long segment hypoplasia" was present from the conduit valve to the main pulmonary artery.The fracture of the previously placed left pulmonary artery stent did not contribute to the need for conduit replacement.The patient tolerated the procedure well with no complications reported.Information related to the balloon valvuloplasty is submitted in medwatch 2025587-2020-02856 g5: corrected the pma number back to p870078 (as was submitted on the initial medwatch).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANCOCK
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key10177064
Report Number2025587-2020-01971
Device Sequence Number19673250
Product Code DYE
UDI-Device Identifier00681490908641
UDI-Public00681490908641
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/12/2014
Device Model NumberHC105-22
Device Catalogue NumberHC105-22
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/22/2020
Supplement Date Received by Manufacturer07/09/2020
08/26/2020
Initial Report FDA Received Date06/19/2020
Supplement Report FDA Received Date07/21/2020
09/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age18 YR
Patient Weight62
-
-