• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TAN/FAN RADILUCENT DROP; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TAN/FAN RADILUCENT DROP; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71631022
Medical Device Problem Codes Malposition of Device (2616); Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Codes Failure of Implant (1924); No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/26/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during trauma surgery - intertan tan hip fracture repair.Targeting is off with guide and drop, therefore screws aren't aligning with nail.There was a delay was under 30 minute however the added time keeps the patient under anesthesia longer than necessary.The procedure was completed using the original device and the surgeon changing their technique- to a freehand approach to target screws.No injury reported.
 
Additional Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection confirmed the device shows significant signs of wear/usage.A functional evaluation of the returned device cannot confirm the stated failure mode.The device will require dimensional evaluation.A dimensional evaluation was conducted and confirms per inspection of the returned complaint device(s), the radiolucent drop is 10 years old and exhibits signs of significant wear/usage - the drill guide is 14 years old and exhibits signs of significant wear/usage.Both parts function as intended.Nothing was detected or determined to be a cause of why the devices are no longer targeting screws correctly.With the parts being as old as they are, it is concluded that the wear/usage on the parts have contributed to the issue stated.A medical investigation was conducted and confirms per subsequent email, the requested relevant clinical information is unavailable.Therefore, a thorough investigation cannot be rendered.Based on the limited information provided, the surgeon¿s change in surgical technique - to a freehand approach to target the screws, and the reported delay of under 30 minutes, did not result in patient injury/impact; therefore, no further medical assessment is warranted at this time.Should additional medical information be provided, this complaint will be re-assessed.A review of complaint history for the listed part revealed no prior complaints for the listed batch.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A review of the risk management file revealed this failure mode was previously identified.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAN/FAN RADILUCENT DROP
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10177291
Report Number1020279-2020-02627
Device Sequence Number2153546
Product Code HSB
UDI-Device Identifier03596010604880
UDI-Public03596010604880
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71631022
Device Catalogue Number71631022
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2020
Initial Date Received by Manufacturer 05/27/2020
Supplement Date Received by Manufacturer05/14/2021
Initial Report FDA Received Date06/21/2020
Supplement Report FDA Received Date05/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-