MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TAN/FAN RADILUCENT DROP; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TAN/FAN RADILUCENT DROP; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71631022
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 05/26/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during trauma surgery - intertan tan hip fracture repair.Targeting is off with guide and drop, therefore screws aren't aligning with nail.There was a delay was under 30 minute however the added time keeps the patient under anesthesia longer than necessary.The procedure was completed using the original device and the surgeon changing their technique- to a freehand approach to target screws.No injury reported.
 
Additional Manufacturer Narrative
Additional information in health effect - impact code and component code.Results of investigation: the device, intended use in treatment, was not returned for evaluation, the reported event could not be confirmed.The clinical/medical team concluded, per subsequent email, the requested relevant clinical information is unavailable.Therefore, a thorough investigation cannot be rendered.Based on the limited information provided, the surgeon¿s change in surgical technique - to a freehand approach to target the screws, and the reported delay of under 30 minutes, did not result in patient injury/impact; therefore, no further medical assessment is warranted at this time.Should additional medical information be provided, this complaint will be re-assessed.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Without the actual product involved, our investigation cannot proceed.Internal complaint reference: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TAN/FAN RADILUCENT DROP
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10177292
Report Number1020279-2020-02628
Device Sequence Number2153547
Product Code HSB
UDI-Device Identifier03596010604880
UDI-Public03596010604880
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K122170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71631022
Device Catalogue Number71631022
Initial Date Received by Manufacturer 05/27/2020
Supplement Date Received by Manufacturer01/05/2021
Initial Report FDA Received Date06/21/2020
Supplement Report FDA Received Date01/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-