Additional information in health effect - impact code and component code.Results of investigation: the device, intended use in treatment, was not returned for evaluation, the reported event could not be confirmed.The clinical/medical team concluded, per subsequent email, the requested relevant clinical information is unavailable.Therefore, a thorough investigation cannot be rendered.Based on the limited information provided, the surgeon¿s change in surgical technique - to a freehand approach to target the screws, and the reported delay of under 30 minutes, did not result in patient injury/impact; therefore, no further medical assessment is warranted at this time.Should additional medical information be provided, this complaint will be re-assessed.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Without the actual product involved, our investigation cannot proceed.Internal complaint reference: (b)(4).
|