The device, used in treatment, was returned for evaluation.A visual inspection of the returned lag driver retaining rod confirms the threaded tip is broken off.The broken piece was not returned.The lag screw driver portion of the device was returned.The device was manufactured in 2015.The device exhibits signs of significant wear and use.A medical investigation will be performed.Per report, the tip broke off into the screw and the screw was extracted.No patient injuries or adverse consequences were reported due to the breakage.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of risk management files found that the reported failure was documented appropriately.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.
|