ZIMMER GMBH DUROM ACETABULAR COMPONENT Back to Search Results
Model Number N/A
Device Problem Insufficient Information (3190)
Patient Problem No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
Concomitant medical product: cls stem hip impl win; item# unknown; lot# unknown.The manufacturer did not receive devices or x-rays for review.As no lot numbers were provided for the devices, the device history records could not be reviewed.Based on an extensive investigation of events reported from several user facilities outside the usa, zimmer identified that the most probable cause for the outcome observed was a loose or unstable cup that resulted from use of surgical techniques not consistent with the manufacturer's recommendations.As a corrective action, a retraining program for users outside the usa was initiated in november 2009 and reported to the national competent authorities as required.The durom cup reported in this case is not marketed in the usa.A similar cup, compatible with the metasul ldh femoral head, is cleared in the usa and a corrective action for this product was reported to the fda in july 2008 as notification z-2415/2426-2008.Since this case is related to the issues for which zimmer implemented a corrective action there will be no further investigation.(b)(4).
 
Event Description
The patient was implanted on the right side and scheduled for a revision surgery on an unknown date due to unknown reasons.
 
Manufacturer Narrative
This follow-up report is being submitted to relay corrected and additional information.Review of event description: patient was scheduled for revision surgery due to unknown reasons.Dhr review: the dhr check could not be performed as the lot number was not available.No further due diligence required as all required information to support the conclusion is available/was already requested.Conclusion: no further investigation required as this issue is known and addressed in cp00000620 (error pattern: potential early revision of the acetabular component due to loosening, implant migration or unresolved pain, higher ion release).At least one of these error patterns is observed in this event.Should any additional information that changes the assessment become available to us, or any extra demand be requested, we will re-evaluate the case.Based on an extensive investigation of events reported from several user facilities outside the usa, zimmer identified that the most probable cause for the outcome observed was a loose or unstable cup that resulted from use of surgical techniques not consistent with the manufacturer's recommendations.As a corrective action, a retraining program for users outside the usa was initiated in november 2009 and reported to the national competent authorities as required.The durom cup reported in this case is not marketed in the usa.A similar cup, compatible with the metasul ldh femoral head, is cleared in the usa and a corrective action for this product was reported to the fda in july 2008 as notification z-2415/2426-2008.Since this case is related to the issues for which zimmer implemented a corrective action there will be no further investigation.Therefore, zimmer gmbh considers this case as closed.Zimmer¿s reference number of this file is (b)(4).
 
Event Description
No change to previously reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUROM ACETABULAR COMPONENT
Type of Device
DUROM ACETABULAR COMPONENT
Manufacturer (Section D)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
MDR Report Key10178475
MDR Text Key195918682
Report Number0009613350-2020-00270
Device Sequence Number1
Product Code KXA
Combination Product (y/n)N
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 09/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/22/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was the Report Sent to FDA? No
Date Manufacturer Received08/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-