| Model Number |
E360T-SY-IN |
| Medical Device Problem Code |
Smoking (1585)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
05/29/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: the field service engineer (fse) evaluated the ventilator and identified that the smoke was from the ekom compressor motor capacitor which the fse then replaced.The ventilator passed all testing per manufacturer¿s specifications and was returned to the customer.If the replaced part is returned for failure investigation, a supplemental medwatch report with the findings will be submitted once the investigation is complete.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a smoke coming from the e360 ventilator compressor.The ventilator was not in use on a patient at the time of the reported event.
|
| |
|