| Model Number |
100149280 |
| Medical Device Problem Code |
Application Program Problem (2880)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/29/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission. .
|
| |
|
Event or Problem Description
|
|
During the procedure, sluggish performance of catheter movement resulting in a 7 second delay was noted.This issue occurred sporadically throughout the procedure resulting in a clinically significant delay in procedure.There no adverse consequences to the patient due to the delay.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, d1, d2, g4, g7, h2, h3, h6, h10.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported sluggish performance and subsequent delay remains unknown, but was resolved after replacing the media converter.
|
| |
|
Event or Problem Description
|
|
During the procedure, sluggish performance of catheter movement was noted with an approximate 7 second delay.This issue occurred sporadically throughout the procedure resulting in a clinically significant delay in procedure.There were no adverse consequences to the patient due to the delay.Further troubleshooting was performed, and once the media converter was replaced it resolved the sluggish performance.
|
| |
|