SMITH & NEPHEW ORTHOPAEDICS AG POLARSTEM COLLAR REAMER GUIDE; PROSTH, HIP, SEMI-CONST, MET/CERAM/POLY, CEMENT OR NON-POROUS, UNCEMENT
|
Back to Search Results |
|
| Model Number |
75102205 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/31/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Primary - dr used the instrument with the hook being broke off but requested it be replaced.No delay & the patient was not affected.Rep is uncertain how it broke.
|
| |
|
Additional Manufacturer Narrative
|
|
A polarstem collar reamer guide, used in treatment, was returned for investigation.The hook of the reamer guide was confirmed to be broken.A design review previously performed demonstrated that the hook may contact the reamer if it is not fully engaged in the rasp during reaming.This is likely to lead to a fracture of the hook and the spring might become loose.It is therefore assumed that foreign material prevented the hook to fully engage in the rasp, which led the reamer to hit and fracture the hook during reaming.A new design of the device has been released in order to prevent the reoccurrence of this issue.The root cause is attributed to a insufficient design.Further investigations have been initiated to review the performance of this device.This complaint has been reopened for reporting reasons, according to the current procedure.No additional information became available.This version of the device will be monitored for similar issues.The returned device will be discarded.
|
| |
|