U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 3.0/26; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO 3.0/26; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 364508
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 05/27/2020
Type of Reportable Event Malfunction
Event or Problem Description
The orsiro drug-eluting stent system was selected to treat a lesion in the rca.The orsiro could not pass the lesion and it was decided to remove it.During removal it was noted that the stent dislodged from the balloon and stayed inside the lesion.A balloon was used to pick-up the dislodged stent and the stent was deployed in the proximal rca.
 
Additional Manufacturer Narrative
The complaint instrument was not returned for a technical investigation.Therefore, no technical investigation on the subject could be performed.The corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.In addition, the angiographic material provided was reviewed.Review of the production documentation for the product detailed above verified that the instrument was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent and the stent retention force of a defined amount of samples is tested from every lot.Review of the angiographic material did not lead to any further information regarding the nature of the complaint.The actual complaint event is not visible.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause could be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO 3.0/26
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key10181324
Report Number1028232-2020-02562
Device Sequence Number3611002
Product Code NIQ
UDI-Device Identifier07640130439177
UDI-Public7640130439177
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/13/2022
Device Model Number364508
Device Catalogue NumberSEE MODEL NO.
Device Lot Number12192524
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/19/2020
Supplement Date Received by Manufacturer07/29/2020
Initial Report FDA Received Date06/22/2020
Supplement Report FDA Received Date07/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-