SMITH & NEPHEW ORTHOPAEDICS AG POLARSTEM COLLAR REAMER GUIDE; PROSTH, HIP, SEMI-CONST, MET/CERAM/POLY, CEMENT OR NON-POROUS, UNCEMENT
|
Back to Search Results |
|
| Model Number |
75102205 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A polarstem collar reamer guide (75102205), used in treatment, was received for evaluation.It was reported that the device failed during the procedure, when the reamer got caught on the reamer guide.It was confirmed, in an additional information request, that the device broke, however no pieces fell inside the wound.A visual analysis could confirm the reported failure mode.The hook and the spring of the device are missing.The rear part of the hook is still attached to the pin of the reamer guide.The production documentation was reviewed, however there were no deviations found.The review of the complaint database showed that similar cases were reported in the past, one of which led to a corrective and preventive action.A design review was performed which demonstrated that the hook may contact the reamer if it is not fully engaged in the rasp during reaming.This is likely to lead to a fracture of the hook and cantilever as reported here.It is therefore assumed that foreign material prevented the hook from fully engaging with the rasp, which led the reamer to hit and fracture the hook intraoperatively.A new design of the device has been released in order to prevent the reoccurrence of this issue.The root cause is attributed to insufficient design specifications.Further actions were started to evaluate the risk of this device.Should the device be returned, this complaint will be reassessed.Smith and nephew will monitor this version of the device for further similar issues.The device will be discarded.
|
| |
|
Event or Problem Description
|
|
It was reported that during inspection the item clip and spring broke apart.No case involved.
|
| |
|
Additional Manufacturer Narrative
|
|
A polarstem collar reamer guide (75102205), used in treatment, was received for evaluation.It was reported that the device failed during the procedure, when the reamer got caught on the reamer guide.It was confirmed, in an additional information request, that the device broke, however no pieces fell inside the wound.A visual analysis could confirm the reported failure mode.The hook and the spring of the device are missing.The rear part of the hook is still attached to the pin of the reamer guide.The production documentation was reviewed, however there were no deviations found.The review of the complaint database showed that similar cases were reported in the past, one of which led to a corrective and preventive action.A design review was performed which demonstrated that the hook may contact the reamer if it is not fully engaged in the rasp during reaming.This collision might lead to a fracture of the hook and the spring can become loose as reported here.It is therefore assumed that foreign material prevented the hook from fully engaging with the rasp, which led the reamer to hit and fracture the hook intraoperatively.A new design of the device has been released in order to prevent the reoccurrence of this issue.The root cause is attributed to insufficient design specifications.Further investigations have been initiated to review the performance of this device.Smith and nephew will monitor this version of the device for further similar issues.The device will be discarded.
|
| |
|
Search Alerts/Recalls
|
|
|