Catalog Number UNKNOWN |
Device Problems
Obstruction of Flow (2423); Device Tipped Over (2589)
|
Patient Problems
Occlusion (1984); Thrombosis (2100)
|
Event Type
Injury
|
Manufacturer Narrative
|
Manufacturers ref# (b)(4).Catalog# is unknown but referred to as cook celect filter.Occupation: non-healthcare professional.Summary of investigational findings: the following allegations have been investigated: acute arterial occlusion.Ivc occlusion/ thrombosis, new dvt, ivc stenosis as a reported complication, is a known risk in relation to filter implant and is well documented in the clinical literature and in clinical practice guidelines.This is supported by the clinical evidence report established to assess available clinical data to identify and evaluate the clinical safety and performance of the cook vena cava filters.Potential adverse events that may occur include, but are not limited to, the following: vena cava occlusion or thrombosis, vena cava stenosis, deep vein thrombosis.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Cook medical will continue to monitor for similar events.
|
|
Event Description
|
Description of event according to initial reporter: it is alleged that "[pt] received a cook celect filter on (b)(6) 2010 for treatment of a deep vein thrombosis ("dvt") and pulmonary embolism.On or around (b)(6) 2018, [pt] was admitted to hospital for an extensive dvt resulting in acute arterial occlusion and requiring an embolectomy and fasciectomy".Patient outcome: it is allege that "[pt] is at risk for future cook ivc filter fractures, migrations, perforations, and tilting.[pt] faces numerous health risks, including the risk of death".
|
|
Event Description
|
Patient is alleging device tilt and dvt post filter placement.Patient notes and further alleges experiencing "extensive dvt (phlegmasia) [status post] filter implant resulting in hospitalization and surgery with embolectomy and fasciotomies, as well as injuries related to tilted ivc filter, including, but not limited to, ongoing medical monitoring, physical limitations, per the (b)(6) 2018 x-ray abdomen: " findings: ivc filter appears tilted right of the midline at the level of l3/l4, unchanged from examination of 2010".
|
|
Manufacturer Narrative
|
Investigation: the following allegations have been investigated: deep vein thrombosis (dvt), tilt, physical limitations.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The additional information regarding deep vein thrombosis (dvt) does not change the previous investigation results for arterial occlusion.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported physical limitations is directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog and lot numbers are unknown, however, the device is manufactured and inspected according to current controls.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Event Description
|
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
|
|
Manufacturer Narrative
|
Manufacturers ref# (b)(4).This report is required by the fda under 21cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Search Alerts/Recalls
|
|
|