MAUDE Adverse Event Report: SMITH & NEPHEW, INC. R3 MULTI HOLE ACETABULAR SHELL 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. R3 MULTI HOLE ACETABULAR SHELL 52MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Model Number 71338665
Medical Device Problem Code Misconnection (1399)
Health Effect - Clinical Code Injury (2348)
Date of Event 05/26/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after surgery the acetabular screw exited the screw hole in the 52mm r3 mulithole cup, this was seen in an x-ray.The surgeon reported the screw was firmly seated at the bottom of the hole with excellent purchase.It appeared to go beyond the cup at the time it was implanted.
 
Event or Problem Description
It was reported that after surgery the acetabular screw exited the screw hole in the 52mm r3 mulithole cup, this was seen in an x-ray.The surgeon reported that both screws were firmly seated at the bottom of the hole with excellent purchase.Neither screw appeared to go beyond the cup at the time they were implanted.
 
Additional Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical/medical investigation concluded that, based on the information provided, the root cause of the reported event could not be definitively concluded.The patient impact beyond the reported screw through the shell could not be determined.No further medical assessment can be rendered at this time.Should additional clinical documentation become available in the future, the clinical/medical task may be re-evaluated.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history did not reveal additional complaints for the listed batch.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors and/or some potential probable causes that could contribute to the reported event have been identified as but not limited to abnormal motion over time, bone degeneration, fit/sizing, lack of ingrowth, lifetime of device, and traumatic injury.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 MULTI HOLE ACETABULAR SHELL 52MM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10183169
Report Number1020279-2020-02641
Device Sequence Number10984714
Product Code MBL
UDI-Device Identifier00885556021491
UDI-Public00885556021491
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K092386
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71338665
Device Catalogue Number71338665
Device Lot Number20CM16369
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/27/2020
Supplement Date Received by Manufacturer06/30/2021
Initial Report FDA Received Date06/23/2020
Supplement Report FDA Received Date07/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-